CHEMVEN

IR of Liquids: Nujol Mull and Thin-Film Spectra

To record IR spectra of liquid organics as thin films and Nujol mulls, recognizing Nujol bands.

Level: UniversitySafety: Needs supervisionTime: 60 min
Aim of Experiment: To record IR spectra of liquid organics as thin films and Nujol mulls, recognizing Nujol bands.

Theoretical Foundations

Liquids are run as thin films between NaCl plates; solids as Nujol (mineral oil) mulls. Nujol contributes its own bands (2925, 1465, 1375 cm⁻¹) which must be recognized and ignored. Comparing film versus mull spectra teaches sampling-artifact recognition.

Required Apparatus & Chemicals

Laboratory Apparatus

  • FTIR spectrometer
  • NaCl plates
  • Spatula
  • Dropper

Chemical Reagents

  • Toluene (liquid)
  • Benzoic acid (solid)
  • Nujol (mineral oil)

Step-by-Step Practical Procedure

  1. Place a drop of toluene between NaCl plates; record 4000-400 cm⁻¹.
  2. Grind benzoic acid with a drop of Nujol; sandwich between plates.
  3. Record the mull spectrum.
  4. Mark the Nujol bands at 2925, 1465, 1375 cm⁻¹.
  5. Assign the remaining bands to the sample.
  6. Compare with a KBr spectrum of the same solid.

Observations & Expected Results

Toluene: aromatic C−H 3030, C=C 1600/1500, 725/695 cm⁻¹. Nujol bands appear in the mull but are absent in KBr spectra.

Worked Stoichiometric Calculations

Problem: A mull shows bands at 2925, 1680 and 1465 cm⁻¹. Which belong to Nujol and which to the sample?

Working: 2925 and 1465 cm⁻¹ are Nujol (C−H stretch/bend); 1680 cm⁻¹ is the sample's C=O stretch.

Safety Precautions & Hazard Control

  • NaCl plates are moisture-sensitive; keep dry.
  • Use minimal Nujol; thick mulls scatter.
  • Clean plates with dry chloroform between samples.
  • Do not use aqueous samples on NaCl plates.

Waste Disposal Protocol

Nujol/organic residues to non-halogenated organic waste.

Viva Voce Oral Defense Questions & Answers

Q: Why must Nujol bands be memorized?

A: They appear in every mull spectrum; mistaking them for sample bands causes misidentification.

Q: Why can't aqueous solutions be run on NaCl plates?

A: Water dissolves the NaCl windows, destroying them.

Q: When is a mull preferred over KBr?

A: For quick qualitative work or moisture-sensitive samples, since mulling is faster and avoids pressing.

← Back to Experiments Directory