Flame AAS: Magnesium in Water with Lanthanum Releasing Agent

To determine magnesium in hard water by flame AAS at 285.2 nm, controlling interferences.

Level: UniversitySafety: Advanced lab onlyTime: 90 min
Aim of Experiment: To determine magnesium in hard water by flame AAS at 285.2 nm, controlling interferences.

Theoretical Foundations

Mg absorbs at 285.2 nm in the air-acetylene flame. Phosphate and aluminium form refractory compounds suppressing Mg atomization; lanthanum chloride (releasing agent) preferentially binds them. Easily ionized Na/K can enhance ionization interference, controlled with an ionization buffer if needed.

Required Apparatus & Chemicals

Laboratory Apparatus

  • Atomic absorption spectrometer
  • Mg hollow-cathode lamp
  • Volumetric flasks
  • Pipettes

Chemical Reagents

  • Mg standard (10 mg L⁻¹)
  • Lanthanum chloride solution
  • KCl ionization buffer (optional)
  • Hard water sample

Step-by-Step Practical Procedure

  1. Add LaCl3 (1% La) to standards and samples.
  2. Prepare Mg standards 0-1 mg L⁻¹ with La.
  3. Install Mg lamp; select 285.2 nm.
  4. Optimize flame; zero with the La blank.
  5. Read standards and samples.
  6. Report Mg in mg L⁻¹.

Observations & Expected Results

Without La, phosphate-rich samples read low; with La the calibration is linear (slope ≈ 0.64 A·L mg⁻¹) and recoveries are quantitative.

Worked Stoichiometric Calculations

Problem: Slope 0.64 A·L mg⁻¹ at 285.2 nm. A sample (with La) gives A = 0.320. Find Mg in mg L⁻¹.

Working: c = 0.320/0.64 = 0.50 mg L⁻¹ Mg.

Safety Precautions & Hazard Control

  • Acetylene safety procedures.
  • Add La to standards and samples equally.
  • Use the lean (oxidizing) flame for Mg.
  • Rinse the burner; La solutions deposit salts.

Waste Disposal Protocol

La/Mg acidic waste to the aqueous heavy-metal waste bottle.

Viva Voce Oral Defense Questions & Answers

Q: What is a chemical interference in AAS?

A: Formation of thermally stable compounds (e.g. Mg3(PO4)2) that do not atomize, lowering the ground-state atom population.

Q: How does lanthanum act as a releasing agent?

A: It competes for the interferent (phosphate), forming LaPO4 and freeing Mg for atomization.

Q: What is ionization interference?

A: At high temperature some Mg ionizes (Mg → Mg⁺ + e⁻); ions do not absorb the atomic line, so easily ionized elements (Na, K) shift the equilibrium and change absorbance.