Flame AAS: Copper in Brass at 324.8 nm

To determine % copper in brass by flame atomic absorption at 324.8 nm.

Level: UniversitySafety: Advanced lab onlyTime: 2 h
Aim of Experiment: To determine % copper in brass by flame atomic absorption at 324.8 nm.

Theoretical Foundations

In AAS, ground-state Cu atoms in an air-acetylene flame absorb the 324.8 nm line from a copper hollow-cathode lamp. Absorbance is proportional to Cu concentration (Beer's law for atoms). AAS is more specific than flame emission because the line source isolates the analyte wavelength.

Required Apparatus & Chemicals

Laboratory Apparatus

  • Atomic absorption spectrometer
  • Cu hollow-cathode lamp
  • Air-acetylene burner
  • Volumetric flasks
  • Pipettes

Chemical Reagents

  • Brass (0.100 g)
  • Nitric acid (conc.)
  • Cu standard (100 mg L⁻¹)
  • Distilled water

Step-by-Step Practical Procedure

  1. Dissolve 0.100 g brass in conc. HNO3 in a fume cupboard; dilute to 100 mL.
  2. Dilute 1.00 mL to 100 mL.
  3. Prepare Cu standards 0-10 mg L⁻¹ in dilute HNO3.
  4. Install the Cu lamp; optimize the 324.8 nm line and flame.
  5. Zero with the acid blank; read standards.
  6. Read the sample; calculate % Cu.

Observations & Expected Results

Absorbance linear with Cu; typical slope ≈ 0.15 A·L mg⁻¹; Zn causes no spectral interference at 324.8 nm.

Worked Stoichiometric Calculations

Problem: Slope 0.15 A·L mg⁻¹. 0.100 g brass to 100 mL; 1.00 mL diluted to 100 mL gives A = 0.975. Find % Cu.

Working: c(measured) = 0.975/0.15 = 6.50 mg L⁻¹. Stock = 650 mg L⁻¹. In 0.100 L: 65.0 mg Cu. %Cu = 65.0/100 × 100 = 65.0%.

Safety Precautions & Hazard Control

  • Acetylene is explosive: leak-check, flashback arrestors, trained supervision.
  • Never look directly at the hollow-cathode beam.
  • Matrix-match standards in dilute HNO3.
  • Allow lamp warm-up for stable output.

Waste Disposal Protocol

Acidic Cu/Zn waste to the aqueous heavy-metal waste bottle.

Viva Voce Oral Defense Questions & Answers

Q: Why is AAS more specific than flame photometry?

A: The hollow-cathode lamp emits only the sharp analyte line, so only ground-state analyte atoms absorb; other elements cannot contribute.

Q: Why use an air-acetylene flame?

A: It is hot enough (~2300 °C) to atomize copper efficiently while keeping ionization low.

Q: What is the role of the hollow-cathode lamp?

A: It provides the intense, narrow resonance line that the analyte atoms absorb.