CHEMVEN

Flame Photometry: Potassium in Fertilizer at 766.5 nm

To determine potassium in fertilizer extract by flame photometry at 766.5 nm.

Level: UniversitySafety: Needs supervisionTime: 60 min
Aim of Experiment: To determine potassium in fertilizer extract by flame photometry at 766.5 nm.

Theoretical Foundations

Potassium emits at 766.5 nm (red region) in the flame, proportional to concentration over 0-10 mg L⁻¹. Fertilizer extracts are diluted into range; results convert to % K2O (× 1.205 from K) for fertilizer grading.

Required Apparatus & Chemicals

Laboratory Apparatus

  • Flame photometer
  • Volumetric flasks
  • Shaker
  • Filter paper
  • Pipettes

Chemical Reagents

  • KCl standard (100 mg L⁻¹ K)
  • Fertilizer sample
  • Distilled water

Step-by-Step Practical Procedure

  1. Extract 1.0 g fertilizer with 200 mL water; filter.
  2. Dilute the extract 1:20.
  3. Prepare K standards 0-10 mg L⁻¹.
  4. Set the K filter (766.5 nm); zero with water.
  5. Read standards and diluted extract.
  6. Apply dilution factor and convert K to % K2O.

Observations & Expected Results

Lilac-red flame emission; linear calibration; typical slope ≈ 6.0 divisions·L mg⁻¹.

Worked Stoichiometric Calculations

Problem: Slope 6.0 divisions·L mg⁻¹. The 1:20 diluted extract reads 30.0 divisions. Find K in the original extract in mg L⁻¹.

Working: c(diluted) = 30.0/6.0 = 5.00 mg L⁻¹. Original = 5.00 × 20 = 100 mg L⁻¹ K.

Safety Precautions & Hazard Control

  • Sodium at high levels overlaps the K filter slightly; correct if needed.
  • Dilute into the linear range.
  • Stabilize the flame.
  • Filter extracts; particles clog the nebulizer.

Waste Disposal Protocol

Dilute salt solutions to the aqueous drain.

Viva Voce Oral Defense Questions & Answers

Q: Why report fertilizer potassium as K2O?

A: Convention: fertilizer grades quote K as the oxide equivalent K2O (factor 1.205).

Q: Why dilute the extract?

A: Fertilizer K is far above the 0-10 mg L⁻¹ linear flame range.

Q: How does sodium interfere?

A: Intense Na emission can bleed through the K filter, giving a positive error at high Na:K ratios.