CHEMVEN

Chemical Deviation from Beer's Law: Dichromate Equilibrium

To demonstrate chemical (non-instrumental) deviation from Beer's law via the chromate-dichromate equilibrium.

Level: UniversitySafety: Needs supervisionTime: 60 min
Aim of Experiment: To demonstrate chemical (non-instrumental) deviation from Beer's law via the chromate-dichromate equilibrium.

Theoretical Foundations

In acid, 2HCrO4⁻ ⇌ Cr2O7²⁻ + H2O. The two species have different spectra and absorptivities, so as total chromium rises the species distribution shifts and A versus c curves downward. This is a true chemical deviation, distinct from stray-light effects.

Required Apparatus & Chemicals

Laboratory Apparatus

  • UV-Vis spectrophotometer
  • Volumetric flasks
  • Pipettes
  • Cuvettes
  • pH meter

Chemical Reagents

  • K2Cr2O7 stock (Cr(VI) solutions 1 × 10⁻⁴ to 5 × 10⁻³ M)
  • Sulfuric acid (dilute)
  • NaOH (dilute)

Step-by-Step Practical Procedure

  1. Prepare Cr(VI) solutions 1 × 10⁻⁴ to 5 × 10⁻³ M at fixed pH 2.
  2. Read A350 for each.
  3. Plot A versus total Cr.
  4. Extrapolate the low-concentration line to high c.
  5. Measure the % deviation at the top concentration.
  6. Repeat one series at pH 8 (pure chromate) to show linearity returns.

Observations & Expected Results

At pH 2 the plot curves downward at high c; at pH 8 (single species CrO4²⁻) it stays linear.

Worked Stoichiometric Calculations

Problem: The low-range line predicts A = 1.50 at the top concentration, but 1.20 is observed. Find the % deviation.

Working: Deviation = (1.20 − 1.50)/1.50 × 100 = −20%.

Safety Precautions & Hazard Control

  • Hold pH rigorously constant; the equilibrium is pH-driven.
  • Distinguish chemical from stray-light curvature (test at pH 8).
  • Cr(VI) is carcinogenic; wear gloves.
  • Use the same wavelength throughout.

Waste Disposal Protocol

Chromium waste to the dedicated bottle.

Viva Voce Oral Defense Questions & Answers

Q: Why is this called a chemical deviation?

A: The absorbing species itself changes with concentration through the dimerization equilibrium, unlike instrumental artifacts.

Q: Why does linearity return at pH 8?

A: Only CrO4²⁻ exists there, so a single absorbing species obeys Beer's law.

Q: How would dilution restore linearity?

A: Dilution shifts the equilibrium toward monomeric HCrO4⁻, making one species dominate again.