Precision Study: Repeatability of Absorbance Readings
To evaluate instrumental precision from replicate absorbance measurements and express it as RSD.
Theoretical Foundations
Precision is quantified by the relative standard deviation (RSD = s/mean × 100%) of n replicate readings of one stable solution. Good UV-Vis instruments give RSD < 1% at mid-range absorbance. This experiment separates random instrument noise from sample preparation error.
Required Apparatus & Chemicals
Laboratory Apparatus
- UV-Vis spectrophotometer
- Volumetric flask
- Pipette
- Cuvette
Chemical Reagents
- KMnO4 solution (~0.5 absorbance units)
- KMnO4 (AR grade, for stock)
- Distilled water (diluent)
Step-by-Step Practical Procedure
- Prepare one KMnO4 solution (A ≈ 0.500 at 525 nm).
- Zero with water; read the same solution 10 times, removing and replacing the cuvette each time.
- Record all 10 absorbances.
- Calculate mean and standard deviation.
- Compute RSD.
- Compare with the <1% criterion.
Observations & Expected Results
Readings cluster tightly, e.g. 0.498-0.503; RSD typically 0.3-0.8%.
Worked Stoichiometric Calculations
Problem: Ten readings give mean 0.500 and SD 0.003. Find the RSD and judge precision.
Working: RSD = 0.003/0.500 × 100 = 0.60%. Below 1%: good precision.
Safety Precautions & Hazard Control
- Use one solution to isolate instrument precision.
- Re-seat the cuvette each time to include positioning error.
- Allow lamp warm-up before starting.
- Avoid touching cuvette optical faces.
Waste Disposal Protocol
Dilute permanganate to oxidizer waste.
Viva Voce Oral Defense Questions & Answers
Q: Precision versus accuracy: distinguish them.
A: Precision is reproducibility (tight clustering); accuracy is closeness to the true value. Good precision does not guarantee accuracy.
Q: Why re-seat the cuvette?
A: To include real-world positioning variability, not just electronic noise.
Q: What dominates noise at very low absorbance?
A: Detector and lamp noise relative to the tiny absorbed signal; hence the 0.1-1.0 working range.