Solvent Cutoff Study: Usable UV Range of Common Solvents
To determine the UV cutoff wavelengths of common solvents and choose correct solvents for UV work.
Theoretical Foundations
Every solvent absorbs below a cutoff wavelength (A = 1.0 for 1 cm): water 190, ethanol 210, hexane 210, chloroform 245, acetone 330 nm. Measurements below the cutoff are meaningless because the solvent, not the sample, absorbs the beam.
Required Apparatus & Chemicals
Laboratory Apparatus
- UV spectrophotometer
- Quartz cuvettes
- Solvent bottles
- Pasteur pipettes
- Wash bottle
Chemical Reagents
- Water
- Ethanol
- Hexane
- Chloroform
- Acetone (HPLC grade)
Step-by-Step Practical Procedure
- Fill a quartz cuvette with each solvent; blank with air.
- Scan 200-400 nm for each solvent.
- Note the wavelength where A reaches 1.0 (cutoff).
- Tabulate cutoffs and compare with literature.
- Decide which solvents suit a 220 nm analysis.
- Repeat for the solvent you plan to use in later experiments.
Observations & Expected Results
Water and hexane transmit to ~200 nm; acetone is opaque below ~330 nm; each solvent shows a sharp absorption edge.
Worked Stoichiometric Calculations
Problem: An acetone blank gives A = 1.5 at 200 nm in a 1 cm cell. Can acetone be used as solvent for a 220 nm determination?
Working: No. Acetone's cutoff is ~330 nm; at 220 nm the solvent absorbs essentially all light, leaving no measurable sample signal.
Safety Precautions & Hazard Control
- Use quartz cuvettes for UV scans.
- Keep cuvettes scrupulously clean; residues absorb in the UV.
- Work in a fume cupboard with volatile solvents.
- Blank with air, not with another solvent.
Waste Disposal Protocol
Return solvents to their bottles or to organic waste as appropriate.
Viva Voce Oral Defense Questions & Answers
Q: What is a solvent cutoff?
A: The wavelength below which the pure solvent itself absorbs too strongly (A > 1) for analytical measurements.
Q: Why is acetone a poor UV solvent?
A: Its n→π* carbonyl absorption sets the cutoff near 330 nm, blocking the whole mid-UV region.
Q: Which solvents suit far-UV work near 200 nm?
A: Water, hexane and cyclohexane, with cutoffs near 190-210 nm.