Peroxidase Activity: Guaiacol Oxidation at 470 nm

To assay peroxidase activity from guaiacol oxidation kinetics at 470 nm.

Level: UniversitySafety: SafeTime: 90 min
Aim of Experiment: To assay peroxidase activity from guaiacol oxidation kinetics at 470 nm.

Theoretical Foundations

Peroxidase catalyses H2O2 oxidation of guaiacol to tetraguaiacol, a brown product (λmax 470 nm, ε = 26.6 mM⁻¹ cm⁻¹). The initial rate dA/dt divided by ε gives enzyme activity in µmol min⁻¹ mL⁻¹ (units per mL).

Required Apparatus & Chemicals

Laboratory Apparatus

  • UV-Vis spectrophotometer
  • Cuvettes
  • Micropipettes
  • Stopwatch
  • Buffer

Chemical Reagents

  • Guaiacol (20 mM)
  • H2O2 (10 mM)
  • Phosphate buffer (pH 7.0)
  • Peroxidase extract (e.g. turnip)

Step-by-Step Practical Procedure

  1. Mix buffer, guaiacol and H2O2 in a cuvette.
  2. Add enzyme extract; start timing immediately.
  3. Record A470 every 15 s for 3 min.
  4. Plot A versus t; take the initial slope.
  5. Convert to activity using ε = 26.6 mM⁻¹ cm⁻¹.
  6. Run a no-enzyme blank.

Observations & Expected Results

Brown colour grows linearly at first; typical ΔA/min ≈ 0.1-0.2 for dilute extracts.

Worked Stoichiometric Calculations

Problem: ΔA470/min = 0.120 (ε = 26.6 mM⁻¹ cm⁻¹, l = 1 cm, 0.1 mL enzyme in 3.0 mL). Find activity in µmol min⁻¹ mL⁻¹.

Working: Rate in cuvette = 0.120/26.6 = 4.51 × 10⁻³ mM min⁻¹ = 4.51 µM min⁻¹. In 3.0 mL: 4.51 × 3.0 = 13.5 nmol min⁻¹ from 0.1 mL enzyme → 135 nmol min⁻¹ mL⁻¹ = 0.135 µmol min⁻¹ mL⁻¹. (Using per-mL-of-reaction-mixture convention: 4.51 µmol L⁻¹ min⁻¹ × 0.003 L/0.0001 L = 0.135 U mL⁻¹.)

Safety Precautions & Hazard Control

  • Start timing at enzyme addition.
  • Use initial linear slope only.
  • Keep H2O2 low; excess inactivates peroxidase.
  • Prepare guaiacol fresh; it darkens in air.

Waste Disposal Protocol

Dilute guaiacol waste to the aqueous drain.

Viva Voce Oral Defense Questions & Answers

Q: What is one enzyme unit here?

A: The amount converting 1 µmol guaiacol per minute under the stated conditions.

Q: Why does excess H2O2 inactivate peroxidase?

A: It over-oxidises the heme centre to inactive forms (suicide inactivation).

Q: Why measure initial rates?

A: Substrate depletion and product inhibition later curve the progress plot.