Standard Addition Method: Iron in a Complex Matrix
To determine iron in a sample with matrix interferences using the standard addition method.
Theoretical Foundations
When the sample matrix alters sensitivity (e.g. phosphate suppressing colour), external calibration fails. In standard addition, known increments of standard are added to equal sample aliquots; plotting A versus added concentration and extrapolating to A = 0 gives the sample concentration, since matrix effects act equally on sample and spikes.
Required Apparatus & Chemicals
Laboratory Apparatus
- UV-Vis spectrophotometer
- Volumetric flasks 100 mL (4)
- Pipettes
- Cuvettes
Chemical Reagents
- Fe standard (10 mg L⁻¹)
- Hydroxylamine hydrochloride
- Acetate buffer
- 1,10-phenanthroline
- Sample with matrix (e.g. phosphate-rich water)
Step-by-Step Practical Procedure
- Add 10.0 mL sample to each of four 100 mL flasks.
- Spike 0, 1.0, 2.0, 3.0 mL of 10 mg L⁻¹ Fe standard.
- Add hydroxylamine, buffer and phenanthroline to each; dilute to mark.
- Wait 15 min; zero at 510 nm with blank.
- Plot A versus added Fe concentration.
- Extrapolate the line to A = 0; the intercept magnitude is the sample concentration in the flask.
Observations & Expected Results
Line rises with slope equal to the in-matrix sensitivity; x-intercept is negative; e.g. absorbances 0.200, 0.280, 0.360, 0.440.
Worked Stoichiometric Calculations
Problem: Added Fe: 0, 0.10, 0.20, 0.30 mg L⁻¹ give A = 0.200, 0.280, 0.360, 0.440. Find the Fe in the diluted flask and in the original sample (10 mL to 100 mL).
Working: Slope = 0.080/0.10 = 0.80 A·L mg⁻¹. x-intercept = −0.200/0.80 = −0.25 mg L⁻¹. Flask conc. = 0.25 mg L⁻¹; original = 0.25 × 10 = 2.50 mg L⁻¹.
Safety Precautions & Hazard Control
- Spikes must be small relative to sample so the matrix is unchanged.
- Use at least three addition levels.
- Ensure the response is linear over the spiked range.
- Identical treatment of all flasks is essential.
Waste Disposal Protocol
Phenanthroline waste to the organic-aqueous waste bottle.
Viva Voce Oral Defense Questions & Answers
Q: When is standard addition preferred over external calibration?
A: When the sample matrix changes the analytical sensitivity, so standards in pure solvent would give a wrong slope.
Q: Why extrapolate to zero absorbance?
A: At A = 0 the added standard exactly cancels the sample's own analyte; the negative intercept magnitude equals the sample concentration.
Q: What is a limitation of standard addition?
A: It corrects proportional (sensitivity) matrix effects but not additive blank interferences, and it needs linearity.