Cholesterol Determination: Liebermann-Burchard Reaction at 620 nm
To determine total serum cholesterol by the Liebermann-Burchard reaction at 620 nm.
Theoretical Foundations
Cholesterol reacts with acetic anhydride and concentrated H2SO4 to give a green-blue polyene chromophore (λmax 620 nm). Serum lipids are first extracted with ethanol/ether or the serum is treated directly with the reagent. Colour develops over ~30 min and is read against cholesterol standards.
Required Apparatus & Chemicals
Laboratory Apparatus
- UV-Vis spectrophotometer
- Test tubes (dry)
- Pipettes
- Stopwatch
- Cuvettes
- Fume cupboard
Chemical Reagents
- Cholesterol standard (200 mg dL⁻¹)
- Acetic anhydride
- Sulfuric acid (conc.)
- Glacial acetic acid
- Serum sample
Step-by-Step Practical Procedure
- Ensure all glassware is completely dry (water destroys the reagent).
- Prepare cholesterol standards 0-300 mg dL⁻¹ in acetic acid.
- Add acetic anhydride then conc. H2SO4 carefully to standards and serum extract.
- Wait 30 min in the dark for full green colour.
- Zero at 620 nm with the reagent blank.
- Read absorbances and determine cholesterol.
Observations & Expected Results
Green-blue colour develops slowly; A620 linear with cholesterol; typical slope ≈ 0.00275 A·dL mg⁻¹.
Worked Stoichiometric Calculations
Problem: Slope 0.00275 A·dL mg⁻¹ at 620 nm. A serum sample gives A = 0.550. Find total cholesterol in mg dL⁻¹.
Working: c = 0.550/0.00275 = 200 mg dL⁻¹ (borderline-high reference range).
Safety Precautions & Hazard Control
- Concentrated H2SO4 + acetic anhydride: extreme corrosive hazard; fume cupboard, goggles, gloves.
- Glassware must be bone dry; water hydrolyses acetic anhydride.
- Develop colour in the dark; light bleaches the chromophore.
- Read at a fixed time; the colour keeps changing.
Waste Disposal Protocol
Strongly acidic anhydride waste: quench carefully, then dispose via corrosive organic waste.
Viva Voce Oral Defense Questions & Answers
Q: Why must everything be dry?
A: Water hydrolyses acetic anhydride to acetic acid, destroying the dehydrating reagent needed for chromophore formation.
Q: Why read at a fixed time?
A: The polyene chromophore forms and then degrades continuously, so absorbance is time-dependent.
Q: What does the reagent do to cholesterol chemically?
A: Strong acid dehydrates and oxidizes cholesterol into conjugated polyene cations that absorb at 620 nm.