Manganese in Steel: Periodate Oxidation Method
To determine manganese in steel by periodate oxidation to permanganate at 525 nm.
Theoretical Foundations
Sodium periodate oxidises Mn(II) to MnO4⁻ in hot nitric acid without a catalyst (λmax 525 nm, ε ≈ 2.3 × 10³). Phosphoric acid masks iron. The method is a standard steelworks analysis for 0.1-2% Mn, with absorbance proportional to manganese after identical oxidation of standards and samples.
Required Apparatus & Chemicals
Laboratory Apparatus
- UV-Vis spectrophotometer
- Hot plate
- Volumetric flasks 100 mL
- Beakers
- Pipettes
- Cuvettes
Chemical Reagents
- Steel drillings (0.200 g)
- Sodium periodate
- Nitric acid (1:1)
- Phosphoric acid
- Mn standard (100 mg L⁻¹)
Step-by-Step Practical Procedure
- Dissolve 0.200 g steel in 20 mL 1:1 HNO3 with heating.
- Add 5 mL H3PO4 and 0.5 g NaIO4; boil 2 min.
- Cool and dilute to 100 mL.
- Oxidize Mn standards (0-2 mg L⁻¹) identically.
- Zero at 525 nm with the reagent blank; read absorbances.
- Calculate % Mn in the steel.
Observations & Expected Results
Purple permanganate develops on boiling; standards and sample treated identically give a linear calibration at 525 nm.
Worked Stoichiometric Calculations
Problem: 0.200 g steel diluted to 100 mL gives A = 0.450 at 525 nm (ε = 2.30 × 10³, l = 1 cm). Find % Mn.
Working: c = 0.450/2.30 × 10³ = 1.96 × 10⁻⁴ mol L⁻¹. Mass in 0.100 L = 1.96 × 10⁻⁴ × 0.100 × 54.94 = 1.07 × 10⁻³ g. %Mn = 1.07 × 10⁻³/0.200 × 100 = 0.538%.
Safety Precautions & Hazard Control
- Dissolve steel completely; undissolved carbides trap manganese.
- Boil with periodate long enough for full oxidation.
- Treat standards through the identical oxidation procedure.
- Avoid chloride; it consumes periodate.
Waste Disposal Protocol
Neutralize acidic waste, reduce permanganate with bisulfite, dispose via heavy-metal waste.
Viva Voce Oral Defense Questions & Answers
Q: Why oxidize the standards as well as the sample?
A: To cancel any incomplete oxidation or matrix effects; both experience identical chemistry, so the calibration is valid.
Q: Why is this a steelworks standard method?
A: It is fast, uses stable permanganate colour, and covers the 0.1-2% Mn range typical of carbon steels.
Q: What does phosphoric acid do here?
A: It forms colourless ferric phosphate, removing the yellow Fe(III) background at 525 nm.