Anthocyanins: pH Differential Method at 520 nm
To determine total monomeric anthocyanins in fruit by the pH differential method at 520 nm.
Theoretical Foundations
Anthocyanins are red flavylium ions at pH 1.0 but convert to colourless hemiketals at pH 4.5. The absorbance difference A = (A520 − A700)pH1 − (A520 − A700)pH4.5 measures only monomeric anthocyanins, excluding polymeric brown pigments. Concentration (as cyanidin-3-glucoside) = A × MW × DF × 1000/(ε × l), with MW = 449.2 and ε = 26900.
Required Apparatus & Chemicals
Laboratory Apparatus
- UV-Vis spectrophotometer
- pH meter
- Volumetric flasks
- Pipettes
- Cuvettes
- Centrifuge
Chemical Reagents
- Fruit/juice sample
- Potassium chloride buffer (pH 1.0)
- Sodium acetate buffer (pH 4.5)
Step-by-Step Practical Procedure
- Dilute the sample identically into pH 1.0 and pH 4.5 buffers (same dilution factor).
- Equilibrate 15 min in the dark.
- Read A520 and A700 of both solutions.
- Compute A = (A520 − A700)pH1 − (A520 − A700)pH4.5.
- Convert to mg L⁻¹ as cyanidin-3-glucoside.
- Compare with literature values for the fruit.
Observations & Expected Results
pH 1.0 solution is bright red, pH 4.5 nearly colourless; the difference absorbance is stable and proportional to monomeric anthocyanin.
Worked Stoichiometric Calculations
Problem: Dilution factor 10. Readings: pH 1.0: A520 = 0.800, A700 = 0.050; pH 4.5: A520 = 0.200, A700 = 0.050. Find total anthocyanins in mg L⁻¹ (MW 449.2, ε 26900).
Working: A = (0.800 − 0.050) − (0.200 − 0.050) = 0.600. c = 0.600 × 449.2 × 10 × 1000/(26900 × 1) = 2,695,200/26,900 = 100.2 mg L⁻¹.
Safety Precautions & Hazard Control
- Use the identical dilution factor for both pH solutions.
- Equilibrate fully; the structural change needs ~15 min.
- Correct each reading at 700 nm for haze.
- Protect from light during equilibration.
Waste Disposal Protocol
Dilute buffer/fruit solutions can go to the aqueous drain.
Viva Voce Oral Defense Questions & Answers
Q: Why measure at two pH values?
A: Only monomeric anthocyanins change colour with pH; polymeric pigments absorb at both pH values and cancel in the difference.
Q: Why subtract A700?
A: To correct for light scattering (haze) by suspended particles, which affects both wavelengths.
Q: Why is the result expressed as cyanidin-3-glucoside?
A: Fruits contain mixtures of anthocyanins; one reference compound with known MW and ε gives a standardized equivalent.