DNA Quantification: A260 Measurement and A260/A280 Purity Ratio

To quantify double-stranded DNA by UV absorbance at 260 nm and assess purity from the A260/A280 ratio.

Level: UniversitySafety: SafeTime: 45 min
Aim of Experiment: To quantify double-stranded DNA by UV absorbance at 260 nm and assess purity from the A260/A280 ratio.

Theoretical Foundations

Nucleic acid bases absorb at 260 nm; for dsDNA, A260 = 1.0 corresponds to 50 µg mL⁻¹ (1 cm path). Proteins absorb at 280 nm, so the A260/A280 ratio indicates purity: ~1.8 for pure DNA, lower when protein-contaminated. Dilution factors convert the measured absorbance to stock concentration.

Required Apparatus & Chemicals

Laboratory Apparatus

  • UV spectrophotometer
  • Quartz cuvettes (micro)
  • Micropipettes
  • TE buffer

Chemical Reagents

  • Purified dsDNA sample (diluted 1:50 in TE buffer)
  • TE buffer (pH 8.0)
  • Nuclease-free water (diluent)
  • RNase A (optional, removes RNA)

Step-by-Step Practical Procedure

  1. Dilute the DNA stock 1:50 in TE buffer.
  2. Blank the instrument with TE buffer from 240-300 nm.
  3. Record A260 and A280 of the diluted sample.
  4. Calculate concentration: c = A260 × 50 (dilution) × 50 µg mL⁻¹.
  5. Compute the A260/A280 ratio.
  6. Judge purity and decide if re-purification is needed.

Observations & Expected Results

Pure DNA shows a smooth peak at 260 nm with A260/A280 ≈ 1.8; protein contamination raises A280 and drops the ratio toward 1.5.

Worked Stoichiometric Calculations

Problem: A 1:50 diluted DNA sample gives A260 = 0.500 and A280 = 0.270. Find the stock concentration and comment on purity.

Working: c = 0.500 × 50 × 50 = 1250 µg mL⁻¹. Ratio = 0.500/0.270 = 1.85 ≈ 1.8, indicating pure DNA.

Safety Precautions & Hazard Control

  • Use quartz cuvettes; glass blocks 260 nm.
  • Blank with the exact buffer the DNA is dissolved in.
  • Keep dilutions accurate; the factor of 50 multiplies any pipetting error.
  • RNA also absorbs at 260 nm; remove it with RNase if DNA-specific values are needed.

Waste Disposal Protocol

Dilute DNA/buffer solutions can go to the aqueous drain.

Viva Voce Oral Defense Questions & Answers

Q: Why is A260 = 1.0 taken as 50 µg mL⁻¹ for dsDNA?

A: It is the empirically established specific absorbance of double-stranded DNA at 260 nm for a 1 cm path.

Q: What does A260/A280 = 1.5 suggest?

A: Protein contamination, since proteins absorb relatively more at 280 nm (aromatic residues) and depress the ratio.

Q: Why must the blank match the sample buffer?

A: Buffer components (EDTA, salts) absorb in the UV; an unmatched blank shifts both A260 and A280.