Sulfate Determination: Turbidimetric Barium Sulfate Method
To determine sulfate in water by turbidimetry of precipitated BaSO4 at 420 nm.
Theoretical Foundations
Ba²⁺ precipitates sulfate as fine BaSO4 crystals that scatter light. The apparent absorbance (turbidity) at 420 nm is proportional to sulfate concentration when crystal size is kept uniform with a conditioning reagent (glycerol-NaCl-HCl). This is nephelometry/turbidimetry rather than true absorption, but Beer's-law-like calibration works.
Required Apparatus & Chemicals
Laboratory Apparatus
- UV-Vis spectrophotometer (turbidity mode)
- Volumetric flasks 100 mL
- Magnetic stirrer
- Cuvettes
- Stopwatch
Chemical Reagents
- Na2SO4 standard (100 mg L⁻¹ SO4²⁻)
- BaCl2 crystals
- Conditioning reagent (glycerol, NaCl, HCl)
- Unknown water sample
Step-by-Step Practical Procedure
- Pipette 0-10 mL sulfate standard into 100 mL flasks (0-10 mg L⁻¹).
- Add 5 mL conditioning reagent to each flask.
- Add a spoonful of BaCl2 crystals, stir exactly 1 min at constant speed.
- Wait 4 min, then read apparent absorbance at 420 nm.
- Treat the unknown identically with identical timing.
- Plot turbidity versus sulfate and read the unknown.
Observations & Expected Results
White turbidity develops; apparent absorbance at 420 nm increases linearly with sulfate; typical slope ≈ 0.020 A·L mg⁻¹.
Worked Stoichiometric Calculations
Problem: Slope is 0.020 A·L mg⁻¹ at 420 nm. A sample gives A = 0.420. Find sulfate in mg L⁻¹.
Working: c = 0.420/0.020 = 21.0 mg L⁻¹ SO4²⁻.
Safety Precautions & Hazard Control
- Stirring time and speed must be identical for every flask; crystal size controls turbidity.
- Read at exactly the same delay after stirring.
- Keep temperature constant; solubility of BaSO4 changes with temperature.
- Filter coloured or turbid samples first.
Waste Disposal Protocol
Collect BaSO4 suspensions; allow to settle and dispose of solids as inorganic waste.
Viva Voce Oral Defense Questions & Answers
Q: Is turbidity a true absorbance? Explain.
A: No: it is light scattered away from the detector by suspended particles, but the instrument reports it as apparent absorbance, which is still proportional to particle concentration.
Q: Why is stirring standardized?
A: Turbidity depends on particle size and number; identical stirring gives reproducible crystal size, hence reproducible scattering.
Q: What is the conditioning reagent for?
A: Glycerol keeps BaSO4 crystals small and suspended, while NaCl/HCl give uniform ionic strength for reproducible precipitation.