Free Chlorine Determination: DPD Method at 515 nm
To determine free residual chlorine in water by the pink DPD oxidation product measured at 515 nm.
Theoretical Foundations
N,N-diethyl-p-phenylenediamine (DPD) is oxidised by free chlorine (HOCl/OCl⁻) to the pink Würster dye (λmax 515 nm, ε ≈ 1.5 × 10⁴ L mol⁻¹ cm⁻¹). At pH 6.2-6.5 only free chlorine reacts rapidly; combined chlorine needs added KI. Absorbance is proportional to free chlorine up to ~2 mg L⁻¹.
Required Apparatus & Chemicals
Laboratory Apparatus
- UV-Vis spectrophotometer
- Volumetric flasks 100 mL
- Pipettes
- Cuvettes
- pH buffer sachets
Chemical Reagents
- Chlorine standard (from standardised bleach)
- DPD reagent tablets or solution
- Phosphate buffer (pH 6.5)
- Unknown chlorinated water
Step-by-Step Practical Procedure
- Buffer 50 mL of each standard and the sample to pH 6.5.
- Add DPD reagent, mix, and start a timer.
- Zero at 515 nm with the buffered blank.
- Read absorbance exactly 2 min after DPD addition.
- Plot the calibration curve (0-2 mg L⁻¹ Cl2).
- Read the unknown free chlorine from the curve.
Observations & Expected Results
Pink colour appears instantly; absorbance at 515 nm linear with free chlorine; colour slowly deepens if reading is delayed.
Worked Stoichiometric Calculations
Problem: A sample gives A = 0.312 at 515 nm (ε = 1.50 × 10⁴, l = 1 cm). Find free chlorine as mg L⁻¹ Cl2.
Working: c = 0.312/1.50 × 10⁴ = 2.08 × 10⁻⁵ mol L⁻¹. mg L⁻¹ = 2.08 × 10⁻⁵ × 70.90 × 1000 = 1.47 mg L⁻¹ Cl2.
Safety Precautions & Hazard Control
- Read at a fixed time (2 min); the colour is not fully stable.
- Control pH at 6.2-6.5; outside this range combined chlorine also reacts.
- Avoid sunlight during colour development.
- Standardise the bleach stock iodometrically the same day.
Waste Disposal Protocol
Dilute DPD solutions can go to the aqueous drain; bulk reagent waste to organic-aqueous waste.
Viva Voce Oral Defense Questions & Answers
Q: Why is pH buffered at 6.5?
A: At pH 6.2-6.5 DPD responds rapidly to free chlorine but only slowly to combined chlorine, giving selectivity.
Q: How would you measure total chlorine with DPD?
A: Add KI: combined chlorine (chloramines) oxidises I⁻ to I2, which then oxidises DPD; the extra colour equals combined chlorine.
Q: Why must the reading time be fixed?
A: The Würster dye slowly continues to form and then fades, so absorbance drifts with time.