Ammonia Determination: Nessler's Reagent Method at 425 nm
To determine ammonia nitrogen in water by the yellow-brown Nessler complex measured at 425 nm.
Theoretical Foundations
Nessler's reagent (K2[HgI4] in KOH) reacts with ammonia to form a yellow-brown colloidal complex (λmax ≈ 425 nm). Rochelle salt is added to prevent precipitation of Ca/Mg hydroxides. Absorbance is proportional to ammonia from ~0.02 to 2 mg L⁻¹ NH3-N.
Required Apparatus & Chemicals
Laboratory Apparatus
- UV-Vis spectrophotometer
- Volumetric flasks 50 mL (7)
- Pipettes
- Cuvettes
- Gloves
Chemical Reagents
- NH4Cl standard (10 mg L⁻¹ as N)
- Nessler's reagent
- Rochelle salt solution
- Unknown water sample
Step-by-Step Practical Procedure
- Pipette 0-5 mL ammonia standard into 50 mL flasks (0-1.0 mg N L⁻¹).
- Add 1 mL Rochelle salt and 1 mL Nessler's reagent to each; dilute to mark.
- Prepare the unknown identically.
- Wait 15 min for colour development; zero at 425 nm with blank.
- Read absorbances promptly (colloid may settle).
- Determine ammonia from the calibration curve.
Observations & Expected Results
Yellow-brown turbidity-like colour develops; absorbance at 425 nm linear with ammonia; typical slope ≈ 0.90 A·L mg⁻¹.
Worked Stoichiometric Calculations
Problem: Slope is 0.90 A·L mg⁻¹ at 425 nm. A sample gives A = 0.450. Find ammonia as mg N L⁻¹.
Working: c = 0.450/0.90 = 0.50 mg N L⁻¹.
Safety Precautions & Hazard Control
- Nessler's reagent contains mercury: wear gloves, avoid skin contact and ingestion.
- Add Rochelle salt to stop Ca/Mg precipitation in hard waters.
- Read promptly; the colloidal product settles with time.
- Distil the sample first if amines or turbidity interfere.
Waste Disposal Protocol
All Nessler waste contains mercury: collect strictly in the mercury waste bottle for specialist disposal.
Viva Voce Oral Defense Questions & Answers
Q: Why is this experiment rated advanced lab only?
A: Nessler's reagent contains toxic mercury compounds requiring strict handling and specialist waste disposal.
Q: What is the purpose of Rochelle salt?
A: It complexes Ca²⁺ and Mg²⁺, preventing their hydroxides from precipitating in the alkaline reagent and scattering light.
Q: Why distil samples before Nesslerization?
A: Distillation separates ammonia from turbidity, colour and interfering amines that would bias the 425 nm reading.