CHEMVEN

Nitrate Determination: Phenoldisulfonic Acid Method at 410 nm

To determine nitrate in water by nitration of phenoldisulfonic acid to a yellow product measured at 410 nm.

Level: UniversitySafety: Needs supervisionTime: 90 min
Aim of Experiment: To determine nitrate in water by nitration of phenoldisulfonic acid to a yellow product measured at 410 nm.

Theoretical Foundations

Nitrate nitrates phenoldisulfonic acid in concentrated sulfuric acid; on making alkaline, the yellow nitrophenoldisulfonate ion forms with λmax ≈ 410 nm. Chloride interferes by forming nitrosyl chloride and must be removed with Ag2SO4. Absorbance at 410 nm follows Beer's law.

Required Apparatus & Chemicals

Laboratory Apparatus

  • UV-Vis spectrophotometer
  • Evaporating dishes
  • Water bath
  • Volumetric flasks 100 mL
  • Pipettes
  • Cuvettes

Chemical Reagents

  • KNO3 standard (100 mg L⁻¹ NO3⁻)
  • Phenoldisulfonic acid reagent
  • Silver sulfate solution
  • Ammonium hydroxide (conc.)
  • Unknown water sample

Step-by-Step Practical Procedure

  1. Remove chloride from sample and standards with Ag2SO4; filter.
  2. Evaporate 25 mL portions to dryness on a water bath.
  3. Add 2 mL phenoldisulfonic acid to each residue; rub with a rod.
  4. Dilute, transfer to 100 mL flasks, make alkaline with NH4OH (yellow develops).
  5. Dilute to mark; zero at 410 nm with blank.
  6. Read absorbances and find nitrate from the calibration curve.

Observations & Expected Results

Yellow colour appears on adding ammonia; absorbance at 410 nm linear with nitrate; typical slope ≈ 0.132 A·L mg⁻¹.

Worked Stoichiometric Calculations

Problem: Slope is 0.132 A·L mg⁻¹ at 410 nm. A sample gives A = 0.330. Find nitrate in mg L⁻¹.

Working: c = 0.330/0.132 = 2.50 mg L⁻¹ NO3⁻.

Safety Precautions & Hazard Control

  • Handle concentrated sulfuric acid reagent with gloves and eye protection.
  • Remove chloride completely; it causes low, erratic results.
  • Evaporate just to dryness; overheating chars organic matter.
  • Add ammonia slowly; the yellow colour needs alkaline pH.

Waste Disposal Protocol

Neutralize acidic residues before disposal in the aqueous waste stream.

Viva Voce Oral Defense Questions & Answers

Q: Why is chloride removed before the determination?

A: Chloride reacts with nitrate in strong acid to give volatile nitrosyl chloride, losing nitrate and depressing results.

Q: Why is the solution made alkaline at the end?

A: The yellow nitrophenoldisulfonate chromophore exists only in alkaline medium; in acid it is colourless.

Q: Why evaporate to dryness?

A: Nitration by phenoldisulfonic acid requires concentrated conditions; water would dilute the reagent and stop the reaction.