Copper Determination: Tetraamminecopper(II) Complex at 610 nm

To determine copper(II) by forming the deep-blue [Cu(NH3)4]²⁺ complex and measuring absorbance at 610 nm.

Level: UniversitySafety: Needs supervisionTime: 60 min
Aim of Experiment: To determine copper(II) by forming the deep-blue [Cu(NH3)4]²⁺ complex and measuring absorbance at 610 nm.

Theoretical Foundations

Cu²⁺ in excess concentrated ammonia forms the deep-blue tetraamminecopper(II) ion with a broad absorption band near 610 nm. With ammonia in large excess the complex dominates the copper speciation, so absorbance at 610 nm is proportional to total copper concentration via a calibration curve.

Required Apparatus & Chemicals

Laboratory Apparatus

  • UV-Vis spectrophotometer
  • Volumetric flasks 100 mL (6)
  • Pipettes
  • Cuvettes
  • Fume cupboard

Chemical Reagents

  • CuSO4 standard solution (100 mg L⁻¹)
  • Concentrated ammonia solution
  • Unknown copper sample
  • Distilled water

Step-by-Step Practical Procedure

  1. Pipette 0-8 mL Cu standard into 100 mL flasks (0-8 mg L⁻¹).
  2. Add 10 mL conc. ammonia to each flask and dilute to mark.
  3. Prepare the unknown identically.
  4. Zero the instrument at 610 nm with the ammonia blank.
  5. Measure absorbances of standards and unknown.
  6. Plot the calibration curve and find the unknown concentration.

Observations & Expected Results

Deep-blue colour intensifies with copper concentration; calibration linear at 610 nm with typical slope ≈ 0.066 A·L mg⁻¹.

Worked Stoichiometric Calculations

Problem: Calibration slope is 0.0663 A·L mg⁻¹ at 610 nm. An unknown gives A = 0.345. Find its copper concentration.

Working: c = A/slope = 0.345/0.0663 = 5.20 mg L⁻¹ Cu.

Safety Precautions & Hazard Control

  • Work in a fume cupboard when handling concentrated ammonia.
  • Keep ammonia concentration identical in all flasks.
  • Use ammonia-resistant stoppers; avoid prolonged skin contact.
  • Measure at the same temperature; complex equilibria shift with temperature.

Waste Disposal Protocol

Collect ammoniacal copper waste in the aqueous heavy-metal waste bottle.

Viva Voce Oral Defense Questions & Answers

Q: Why must ammonia be in large excess?

A: To convert essentially all Cu²⁺ into [Cu(NH3)4]²⁺ so absorbance reflects total copper, not a mixture of species.

Q: Why is 610 nm chosen?

A: It is the absorption maximum of the tetraammine complex, giving maximum sensitivity and minimum wavelength error.

Q: Which ions interfere in this method?

A: Ni²⁺ (blue complex) and high Fe³⁺; also ions precipitated by ammonia such as Al³⁺ scatter light.