Manganese Determination: Persulfate Oxidation to Permanganate
To determine manganese in steel by oxidizing Mn(II) to permanganate with ammonium persulfate and measuring at 525 nm.
Theoretical Foundations
Mn(II) is oxidised by persulfate, catalysed by Ag⁺, to the intensely purple MnO4⁻ ion (λmax 525 nm, ε ≈ 2.3 × 10³ L mol⁻¹ cm⁻¹). Phosphoric acid masks Fe(III) as a colourless complex so it does not interfere. Absorbance of the resulting permanganate is proportional to the original manganese.
Required Apparatus & Chemicals
Laboratory Apparatus
- UV-Vis spectrophotometer
- Volumetric flasks 250 mL and 100 mL
- Hot plate
- Beakers
- Pipettes
- Cuvettes
Chemical Reagents
- Ammonium persulfate (solid)
- Silver nitrate solution (0.1 M)
- Phosphoric acid (conc.)
- Mn standard solution (50 mg L⁻¹)
- Steel sample
- Nitric acid for dissolution
Step-by-Step Practical Procedure
- Dissolve 0.500 g steel in dilute HNO3 and dilute to 250 mL.
- Prepare Mn standards 0-4 mg L⁻¹ in 100 mL flasks.
- Add 5 mL H3PO4, 2 mL AgNO3 and 1 g persulfate to each flask and the sample aliquot.
- Heat to boiling 2 min until the purple colour fully develops; cool.
- Dilute to mark, zero at 525 nm with a reagent blank.
- Read absorbances and determine Mn from the calibration curve.
Observations & Expected Results
Purple permanganate colour develops on heating; calibration linear at 525 nm; phosphoric acid keeps the solution free of yellow Fe(III) colour.
Worked Stoichiometric Calculations
Problem: 0.500 g steel dissolved to 250 mL gives A = 0.380 at 525 nm (ε = 2.30 × 10³, l = 1 cm). Calculate % Mn in the steel.
Working: c = 0.380/2.30 × 10³ = 1.65 × 10⁻⁴ mol L⁻¹. Mass Mn in 0.250 L = 1.65 × 10⁻⁴ × 0.250 × 54.94 = 2.27 × 10⁻³ g. %Mn = 2.27 × 10⁻³/0.500 × 100 = 0.454%.
Safety Precautions & Hazard Control
- Boil long enough for complete oxidation; incomplete oxidation gives low results.
- Cool before diluting to the mark (volume changes with temperature).
- Do not overheat after colour develops; permanganate decomposes on prolonged boiling.
- Keep chloride out; Ag⁺ precipitates AgCl and loses catalytic action.
Waste Disposal Protocol
Neutralize acidic permanganate waste, reduce with bisulfite, and dispose via the heavy-metal aqueous waste stream.
Viva Voce Oral Defense Questions & Answers
Q: What is the role of Ag⁺ in this method?
A: Ag⁺ catalyses the oxidation of Mn(II) to MnO4⁻ by persulfate, which is otherwise very slow.
Q: Why is phosphoric acid added?
A: It complexes Fe(III) as colourless [Fe(PO4)2]³⁻, preventing the yellow iron colour from interfering at 525 nm.
Q: Why must chloride be absent?
A: Chloride precipitates the Ag⁺ catalyst as AgCl and can also be oxidised, consuming persulfate.