Constant Current versus Controlled Potential Deposition
Show that controlled-potential deposition separates copper from nickel, while constant current does not.
Theoretical Foundations
Cu2+/Cu (+0.34 V) and Ni2+/Ni (−0.25 V) have well-separated reduction potentials. At constant current the cathode potential drifts negative as copper depletes, eventually depositing nickel too. With a potentiostat holding the cathode at −0.30 V vs SCE, only copper deposits (Cu2+ + 2e- → Cu) while Ni2+ stays in solution — demonstrating selective electrogravimetry.
Required Apparatus & Chemicals
Laboratory Apparatus
- Analytical balance
- Platinum electrodes
- Potentiostat with reference electrode (SCE)
- DC power supply
- Two beakers
- Oven
- Desiccator
Chemical Reagents
- Cu-Ni standard solution (50.0% Cu)
- Sulfuric acid, dilute
- Distilled water
- Ethanol
Step-by-Step Practical Procedure
- Place identical aliquots of Cu-Ni solution in two cells; weigh two cathodes.
- Cell A: electrolyse at constant 2 A for 45 minutes.
- Cell B: electrolyse at controlled −0.30 V vs SCE until the current falls to background.
- Wash (current/potential maintained), dry and weigh both cathodes.
- Calculate apparent % Cu for each and compare with the true 50.0%.
Observations & Expected Results
Deposit A is slightly off-colour (nickel codeposited); deposit B is pure pink copper.
Worked Stoichiometric Calculations
Problem: 0.5000 g alloy equivalent per aliquot. Constant current gave 0.3015 g deposit; controlled potential gave 0.2490 g. Calculate apparent % Cu for each (true = 50.0%).
Working: Constant current: (0.3015/0.5000) × 100 = 60.3% (high — nickel codeposited). Controlled potential: (0.2490/0.5000) × 100 = 49.8% ≈ true value. Potential control achieves the separation.
Safety Precautions & Hazard Control
- Position the reference electrode close to the cathode.
- Let the controlled-potential run reach background current for completeness.
- Do not let the constant-current cell run so long that the deposit burns.
Waste Disposal Protocol
Electrolytes to heavy-metal waste.
Viva Voce Oral Defense Questions & Answers
Q: Why does constant current eventually deposit nickel?
A: As Cu2+ depletes, maintaining the fixed current forces the cathode potential more negative until Ni2+ reduction can sustain it — selectivity is lost.
Q: What is the role of the reference electrode?
A: It lets the potentiostat hold the working electrode at a fixed potential, so only species reducible at that potential deposit.