Nickel in Laterite Ore as Ni(DMG)2
Determine low-grade nickel in a laterite ore by DMG gravimetry after iron removal.
Theoretical Foundations
Laterites contain 1-3% Ni in a matrix of iron oxides. The ore is dissolved in acid; iron is removed by precipitating Fe(OH)3 with ammonia (nickel stays soluble as the ammine at controlled pH, or is separated by double precipitation). Nickel is then precipitated: Ni2+ + 2 H2DMG → Ni(HDMG)2(s) + 2H+, GF = 0.2032. A 1 g sample keeps the precipitate weighable despite the low grade.
Required Apparatus & Chemicals
Laboratory Apparatus
- Analytical balance
- 250 mL beakers
- Hot plate
- Quantitative filter paper
- Sintered-glass crucible
- Oven
- Desiccator
Chemical Reagents
- Laterite ore, finely ground
- Hydrochloric acid, concentrated and dilute
- Ammonia solution
- Dimethylglyoxime solution, 1% in ethanol
- Distilled water
Step-by-Step Practical Procedure
- Weigh accurately 1.0 g of ground ore.
- Digest with 20 mL conc. HCl on a hot plate until decomposed; evaporate to low volume.
- Dilute to 150 mL, add ammonia to precipitate Fe(OH)3 at pH ~4 (methyl red endpoint); filter off iron.
- To the filtrate add DMG solution at 70-80 C, then ammonia to pH 8-9.
- Digest, filter the scarlet precipitate through a weighed sintered-glass crucible.
- Wash, dry at 150 C to constant mass and weigh; calculate % Ni.
Observations & Expected Results
Bulky brown Fe(OH)3 is filtered off; the pale filtrate gives a small scarlet precipitate, typically 0.05-0.10 g.
Worked Stoichiometric Calculations
Problem: 1.000 g of laterite gave 0.0725 g Ni(DMG)2. Calculate % Ni in the ore.
Working: Mass Ni = 0.0725 × 0.2032 = 0.01473 g. % Ni = (0.01473/1.000) × 100 = 1.47%.
Safety Precautions & Hazard Control
- Remove iron quantitatively or it contaminates the nickel precipitate.
- Grind the ore finely for complete acid attack.
- Work in a fume cupboard during acid digestion.
Waste Disposal Protocol
Acidic iron/nickel residues to heavy-metal waste.
Viva Voce Oral Defense Questions & Answers
Q: Why is iron removed before the nickel precipitation?
A: Fe3+ would precipitate as Fe(OH)3 when ammonia is added for the DMG step, massively contaminating the small nickel precipitate.
Q: Why take a full 1 g sample?
A: At 1-2% Ni the precipitate from a smaller sample would be too light to weigh accurately; a larger sample gives a weighable mass.