Coulometric Titration Concept: Acid Determination (Dry-Lab)

To determine an acid's concentration from a coulometric titration as a dry-lab exercise using Faraday's law.

Level: UniversitySafety: SafeTime: 45 min
Aim of Experiment: To determine an acid's concentration from a coulometric titration as a dry-lab exercise using Faraday's law.

Theoretical Foundations

In coulometry the titrant (OH-) is generated electrolytically at 100% current efficiency: cathode reaction 2 H2O + 2 e- -> H2 + 2 OH-. Moles OH- = Q/F = I t/F. The endpoint is detected potentiometrically or with an indicator. No standard titrant solution is needed - the coulomb is the primary standard.

Required Apparatus & Chemicals

Laboratory Apparatus

  • (Dry-lab) Coulometric data: constant current, time to endpoint
  • Constant-current source (concept)
  • Platinum generating electrodes (concept)
  • Timer (concept)

Chemical Reagents

  • (Dry-lab) HCl sample
  • KBr supporting electrolyte
  • Indicator (concept)

Step-by-Step Practical Procedure

  1. Note the constant current I and the time t to the endpoint.
  2. Compute Q = I t.
  3. Compute moles of OH- generated = Q/F.
  4. Equate to moles of acid titrated.
  5. Calculate the acid molarity.
  6. List the requirements for valid coulometry.

Observations & Expected Results

A 25 mL sample of ~0.01 M acid needs ~240 C at 0.1 A (~40 min); computed molarity matches the prepared value.

Worked Stoichiometric Calculations

Problem: 25.0 mL HCl titrated at 0.100 A for 241 s to the endpoint. Find [HCl].

Working: Q = 0.100 x 241 = 24.1 C. Moles e- = 24.1/96485 = 2.498 x 10^-4 mol = moles OH- = moles HCl. [HCl] = 2.498 x 10^-4/0.0250 = 9.99 x 10^-3 M.

Safety Precautions & Hazard Control

  • Dry-lab only.
  • Real coulometry demands 100% current efficiency - no side reactions at the generating electrode.
  • The endpoint detection must be sharp.

Waste Disposal Protocol

No wet waste.

Viva Voce Oral Defense Questions & Answers

Q: Why is the coulomb called a primary standard?

A: Charge is measured absolutely (current x time) with no chemical standardisation needed, provided current efficiency is 100%.

Q: What could reduce current efficiency?

A: Side electrode reactions (e.g. O2 evolution, impurity reduction) that consume charge without generating titrant.