Demineralisation of Water: Mixed-Bed Ion Exchange
To produce demineralised water with a mixed-bed (cation + anion) exchange column and check its conductivity.
Theoretical Foundations
A mixed bed contains H+-form cation resin and OH--form anion resin intimately mixed. Cations release H+, anions release OH-, and the two combine: H+ + OH- -> H2O. The effluent conductivity should fall below ~1 uS/cm (distilled-water quality). Regeneration separates the resins by backwashing (density difference) before acid/caustic treatment.
Required Apparatus & Chemicals
Laboratory Apparatus
- Mixed-bed column
- Conductivity meter
- Measuring cylinder
- Beakers
Chemical Reagents
- Tap water
- Distilled water (for comparison)
- pH test strips (bed-exhaustion check)
- Regeneration chemicals (instructor use only)
Step-by-Step Practical Procedure
- Rinse the mixed-bed column with distilled water.
- Pass tap water through at ~5 mL/min, discarding the first 50 mL.
- Collect 200 mL of effluent.
- Measure its conductivity and compare with tap water and distilled water.
- Test for residual chloride (AgNO3) and hardness (EDTA or soap test).
- Calculate the % reduction in conductivity.
Observations & Expected Results
Tap water (~300-600 uS/cm) drops to <5 uS/cm; chloride and hardness tests become negative.
Worked Stoichiometric Calculations
Problem: Tap water 480 uS/cm; effluent 3.2 uS/cm. What is the % demineralisation by conductivity?
Working: % removal = (480 - 3.2)/480 x 100 = 99.3%.
Safety Precautions & Hazard Control
- Do not exceed the recommended flow rate; channelling ruins quality.
- Discard the first effluent (fines and air).
- Keep the bed wet between uses.
Waste Disposal Protocol
Water effluents to drain.
Viva Voce Oral Defense Questions & Answers
Q: Why does a mixed bed give purer water than two separate beds?
A: H+ and OH- released are instantly consumed as water, driving both exchanges to completion with no residual counter-ions leaking through.
Q: Why discard the first 50 mL?
A: It contains resin fines, air bubbles and the rinse-water front, not representative product water.