Ion Chromatography Concept: Anions in Drinking Water (Dry-Lab)

To quantify fluoride, chloride, nitrate and sulfate in drinking water from an ion chromatogram as a dry-lab exercise.

Level: UniversitySafety: SafeTime: 60 min
Aim of Experiment: To quantify fluoride, chloride, nitrate and sulfate in drinking water from an ion chromatogram as a dry-lab exercise.

Theoretical Foundations

Ion chromatography separates anions on an anion-exchange column with carbonate/bicarbonate eluent; a suppressor converts the eluent to weakly conducting carbonic acid while converting analyte anions to highly conducting acids, so conductivity detection is sensitive. External calibration converts peak areas to mg/L; results are checked against WHO guidelines.

Required Apparatus & Chemicals

Laboratory Apparatus

  • (Dry-lab) Ion chromatogram + calibration data
  • Calculator
  • Suppressor unit diagram (concept)
  • Calibration data sheet
  • WHO guideline table

Chemical Reagents

  • (Dry-lab) Carbonate eluent composition table
  • (Dry-lab) Anion standard data set
  • (Dry-lab exercise: sample chromatogram)

Step-by-Step Practical Procedure

  1. Identify each peak by retention time versus standards.
  2. Note the elution order: F- < Cl- < NO2- < NO3- < SO4(2-).
  3. Build calibration lines (area vs mg/L) from standard data.
  4. Read sample peak areas and compute each anion's mg/L.
  5. Compare with WHO guidelines (F- 1.5, Cl- 250, NO3- 50, SO4(2-) 250 mg/L).
  6. Explain the suppressor's role.

Observations & Expected Results

Typical drinking water: F- 0.5, Cl- 30, NO3- 10, SO4(2-) 40 mg/L - all within guidelines.

Worked Stoichiometric Calculations

Problem: Nitrate calibration: area = 520 x C (mg/L). Sample nitrate area = 7800. Find mg/L NO3-.

Working: C = 7800/520 = 15.0 mg/L - below the 50 mg/L WHO guideline.

Safety Precautions & Hazard Control

  • Dry-lab exercise.
  • In practice, filter samples (0.45 um) and use a guard column.
  • Degas the eluent to protect the suppressor.

Waste Disposal Protocol

No wet waste in the dry-lab version.

Viva Voce Oral Defense Questions & Answers

Q: What does the suppressor do?

A: It exchanges Na+ for H+ after the column: the carbonate eluent becomes weakly conducting carbonic acid while analyte anions become highly conducting acids, boosting sensitivity.

Q: Why does sulfate elute last?

A: Its double charge binds it most strongly to the anion-exchange stationary phase.