Ion Chromatography Concept: Anions in Drinking Water (Dry-Lab)
To quantify fluoride, chloride, nitrate and sulfate in drinking water from an ion chromatogram as a dry-lab exercise.
Theoretical Foundations
Ion chromatography separates anions on an anion-exchange column with carbonate/bicarbonate eluent; a suppressor converts the eluent to weakly conducting carbonic acid while converting analyte anions to highly conducting acids, so conductivity detection is sensitive. External calibration converts peak areas to mg/L; results are checked against WHO guidelines.
Required Apparatus & Chemicals
Laboratory Apparatus
- (Dry-lab) Ion chromatogram + calibration data
- Calculator
- Suppressor unit diagram (concept)
- Calibration data sheet
- WHO guideline table
Chemical Reagents
- (Dry-lab) Carbonate eluent composition table
- (Dry-lab) Anion standard data set
- (Dry-lab exercise: sample chromatogram)
Step-by-Step Practical Procedure
- Identify each peak by retention time versus standards.
- Note the elution order: F- < Cl- < NO2- < NO3- < SO4(2-).
- Build calibration lines (area vs mg/L) from standard data.
- Read sample peak areas and compute each anion's mg/L.
- Compare with WHO guidelines (F- 1.5, Cl- 250, NO3- 50, SO4(2-) 250 mg/L).
- Explain the suppressor's role.
Observations & Expected Results
Typical drinking water: F- 0.5, Cl- 30, NO3- 10, SO4(2-) 40 mg/L - all within guidelines.
Worked Stoichiometric Calculations
Problem: Nitrate calibration: area = 520 x C (mg/L). Sample nitrate area = 7800. Find mg/L NO3-.
Working: C = 7800/520 = 15.0 mg/L - below the 50 mg/L WHO guideline.
Safety Precautions & Hazard Control
- Dry-lab exercise.
- In practice, filter samples (0.45 um) and use a guard column.
- Degas the eluent to protect the suppressor.
Waste Disposal Protocol
No wet waste in the dry-lab version.
Viva Voce Oral Defense Questions & Answers
Q: What does the suppressor do?
A: It exchanges Na+ for H+ after the column: the carbonate eluent becomes weakly conducting carbonic acid while analyte anions become highly conducting acids, boosting sensitivity.
Q: Why does sulfate elute last?
A: Its double charge binds it most strongly to the anion-exchange stationary phase.