CHEMVEN

Van Deemter Curve Concept: Optimising HPLC Flow Rate (Dry-Lab)

To find the optimum HPLC flow rate from a van Deemter plot as a dry-lab exercise.

Level: UniversitySafety: SafeTime: 60 min
Aim of Experiment: To find the optimum HPLC flow rate from a van Deemter plot as a dry-lab exercise.

Theoretical Foundations

The van Deemter equation H = A + B/u + C u describes plate height versus linear velocity: eddy diffusion (A, constant), longitudinal diffusion (B/u, dominates at low flow), and mass-transfer resistance (C u, dominates at high flow). The minimum H (maximum efficiency) occurs at u_opt = sqrt(B/C). Operating near u_opt gives the best resolution per unit time.

Required Apparatus & Chemicals

Laboratory Apparatus

  • (Dry-lab) H vs u data table
  • Graph paper / spreadsheet
  • Curve-fitting guide
  • Column diameter specification

Chemical Reagents

  • (Dry-lab exercise: van Deemter data set)
  • (Dry-lab exercise: worked fitting example)
  • (Dry-lab exercise: particle-size reference table)

Step-by-Step Practical Procedure

  1. Plot H versus u from the data.
  2. Locate the minimum visually.
  3. Fit H = A + B/u + C u (or read A, B, C from the fit).
  4. Calculate u_opt = sqrt(B/C) and H_min.
  5. Convert u_opt to a volumetric flow rate for the given column diameter.
  6. Discuss the trade-off of running above u_opt.

Observations & Expected Results

The curve is U-shaped with a minimum around 1 mm/s for 5 um particles; H_min ~10-15 um.

Worked Stoichiometric Calculations

Problem: Fit gives A = 4.0 um, B = 8.0 um.mm/s, C = 2.0 um.s/mm. Find u_opt and H_min.

Working: u_opt = sqrt(B/C) = sqrt(8.0/2.0) = 2.0 mm/s. H_min = A + 2 sqrt(BC) = 4.0 + 2 sqrt(16) = 12.0 um.

Safety Precautions & Hazard Control

  • Dry-lab exercise.
  • Note the equation form: some texts use reduced parameters.
  • Real optimisation also weighs analysis time and pressure limits.

Waste Disposal Protocol

No waste.

Viva Voce Oral Defense Questions & Answers

Q: Explain each van Deemter term.

A: A: eddy diffusion from multiple path lengths (flow-independent); B/u: longitudinal molecular diffusion (worse at low flow); C u: resistance to mass transfer between phases (worse at high flow).

Q: Why not always run at u_opt?

A: It may be slow; many methods run 2-3x faster accepting slightly higher H to save time, within pressure limits.

← Back to Experiments Directory