Cyclic Voltammetry Concept: Reversibility of Fe(CN)6(3-/4-) (Dry-Lab)

To diagnose reversibility of the ferro/ferricyanide couple from a cyclic voltammogram as a dry-lab exercise.

Level: UniversitySafety: SafeTime: 60 min
Aim of Experiment: To diagnose reversibility of the ferro/ferricyanide couple from a cyclic voltammogram as a dry-lab exercise.

Theoretical Foundations

For a reversible one-electron couple at 25 C, the peak separation delta_Ep = Epa - Epc = 59 mV, ipa/ipc = 1, and peak current follows Randles-Sevcik: ip = 2.69 x 10^5 n^3/2 A D^1/2 C v^1/2. Larger separations or scan-rate-dependent shifts indicate quasi-reversible or irreversible behaviour. Ep is independent of scan rate for reversible systems.

Required Apparatus & Chemicals

Laboratory Apparatus

  • (Dry-lab) CVs of 5 mM K3Fe(CN)6 at several scan rates
  • Ruler
  • Calculator
  • Nitrogen purge setup (concept)
  • Polishing kit for glassy carbon (concept)

Chemical Reagents

  • (Dry-lab) K3Fe(CN)6
  • KCl supporting electrolyte
  • Glassy carbon electrode (concept)

Step-by-Step Practical Procedure

  1. Measure Epa and Epc at each scan rate from the voltammograms.
  2. Compute delta_Ep and check against 59 mV.
  3. Measure ipa and ipc; compute the ratio.
  4. Plot ip versus sqrt(scan rate) and check linearity.
  5. Use the Randles-Sevcik slope with given A to estimate D.
  6. Conclude whether the couple is reversible.

Observations & Expected Results

delta_Ep ~60-70 mV, ipa/ipc ~1.0, and ip linear in v^1/2 - the textbook reversible case.

Worked Stoichiometric Calculations

Problem: Slope of ip vs v^1/2 = 3.20 x 10^-4 A/(V/s)^1/2; A = 0.0707 cm2, C = 5.0 mM, n = 1. Estimate D.

Working: Randles-Sevcik: slope = 2.69 x 10^5 n^3/2 A D^1/2 C. With C = 5.0 x 10^-6 mol/cm3: 3.20 x 10^-4 = 2.69 x 10^5 x 0.0707 x D^1/2 x 5.0 x 10^-6 -> D^1/2 = 3.366 x 10^-3; D = 1.13 x 10^-5 cm2/s.

Safety Precautions & Hazard Control

  • Dry-lab only; real CV needs deaerated cells and polished electrodes.
  • In practice, polish the working electrode before each run.
  • Keep the reference electrode's filling solution topped up.

Waste Disposal Protocol

No wet waste in the dry-lab version.

Viva Voce Oral Defense Questions & Answers

Q: State the three reversibility criteria.

A: delta_Ep = 59/n mV at 25 C, ipa/ipc = 1, and Ep independent of scan rate (with ip proportional to v^1/2).

Q: What does a larger delta_Ep mean?

A: Slower electron transfer: quasi-reversible or irreversible kinetics.

Q: Why plot ip vs sqrt(v)?

A: Diffusion-controlled peak current scales with v^1/2 (Randles-Sevcik); linearity confirms diffusion control.