Calcium ISE: Water Hardness Screening
To screen calcium in drinking-water samples with a calcium ion-selective electrode and calibration curve.
Theoretical Foundations
The calcium ISE (PVC membrane with ionophore) follows E = const + (0.05916/2) log[Ca2+] = const + 0.02958 log[Ca2+] at 25 C. An ionic strength adjuster keeps activity coefficients constant so the electrode reads concentration. Results in mg/L Ca2+ convert to hardness contribution: mg/L as CaCO3 = 2.497 x mg/L Ca2+.
Required Apparatus & Chemicals
Laboratory Apparatus
- Calcium ISE + reference
- mV meter
- Volumetric flasks
- Beakers
Chemical Reagents
- CaCl2 standard (1000 mg/L Ca2+)
- Ionic strength adjuster (KCl/ISA)
- Water samples
- Distilled water
Step-by-Step Practical Procedure
- Prepare Ca2+ standards: 1, 10, 50, 100 mg/L with ISA added.
- Measure mV of each standard with stirring.
- Plot E vs log[Ca2+]; check slope (~+29.6 mV/decade).
- Measure water samples with the same ISA proportion.
- Read [Ca2+] from the calibration line.
- Convert to mg/L as CaCO3 and classify hardness (soft/moderate/hard).
Observations & Expected Results
Slope is typically +28 to +30 mV/decade; tap water often reads 20-80 mg/L Ca2+ (50-200 mg/L as CaCO3).
Worked Stoichiometric Calculations
Problem: Sample reads 60 mg/L Ca2+. Express as mg/L CaCO3 and classify.
Working: As CaCO3 = 60 x 2.497 = 150 mg/L - moderately hard (61-120 soft-moderate boundary varies; 120-180 is hard by USGS).
Safety Precautions & Hazard Control
- Add ISA to every solution including samples.
- Condition the electrode per the manual; do not let the membrane dry.
- Rinse between standards from low to high concentration.
Waste Disposal Protocol
Calcium standards to drain.
Viva Voce Oral Defense Questions & Answers
Q: Why is the slope ~29.6 mV/decade?
A: The Nernst factor for a divalent ion is 0.05916/2 = 0.02958 V per decade at 25 C.
Q: Why add an ionic strength adjuster?
A: The electrode senses activity; constant ionic strength makes activity proportional to concentration so calibration holds.