CHEMVEN

Oxygen Electrode Concept: Dissolved Oxygen by Winkler vs Probe

To determine dissolved oxygen in water by the Winkler titration and compare with a Clark oxygen-electrode reading.

Level: UniversitySafety: Needs supervisionTime: 90 min
Aim of Experiment: To determine dissolved oxygen in water by the Winkler titration and compare with a Clark oxygen-electrode reading.

Theoretical Foundations

Winkler: Mn2+ + 2 OH- -> Mn(OH)2; 2 Mn(OH)2 + O2 -> 2 MnO(OH)2; on acidification with I-, MnO(OH)2 + 2 I- + 4 H+ -> Mn2+ + I2 + 3 H2O; I2 is titrated with thiosulfate (1 mL 0.025 M Na2S2O3 = 0.2 mg O2). The Clark probe measures O2 amperometrically through a membrane; comparing the two validates the probe.

Required Apparatus & Chemicals

Laboratory Apparatus

  • BOD bottles
  • Burette
  • Clark dissolved-oxygen probe + meter
  • Pipettes

Chemical Reagents

  • MnSO4 solution
  • Alkaline KI reagent
  • Conc. H2SO4
  • Na2S2O3 standard (0.025 M)
  • Starch indicator

Step-by-Step Practical Procedure

  1. Fill a BOD bottle completely with the sample, avoiding air bubbles.
  2. Add MnSO4 then alkaline KI below the surface; stopper and mix.
  3. Let the brown precipitate settle; add conc. H2SO4 and mix to dissolve.
  4. Titrate 100 mL of the iodine solution with 0.025 M thiosulfate to a starch endpoint.
  5. Calculate mg/L dissolved oxygen.
  6. Measure the same water with the calibrated Clark probe.
  7. Compare the two results.

Observations & Expected Results

Winkler titres give e.g. 8.0 mL thiosulfate per 100 mL sample; the probe typically agrees within 0.3 mg/L.

Worked Stoichiometric Calculations

Problem: 100 mL sample needed 8.20 mL of 0.0250 M Na2S2O3. Calculate dissolved O2 in mg/L.

Working: 1 mL 0.025 M thiosulfate = 0.2 mg O2. O2 = 8.20 x 0.2 = 1.64 mg per 100 mL = 16.4 mg/L. (High value indicates aerated/supersaturated sample; typical surface water is 7-9 mg/L.)

Safety Precautions & Hazard Control

  • Exclude all air bubbles when stoppering the BOD bottle.
  • Add reagents below the liquid surface.
  • Acidify carefully; conc. H2SO4 is corrosive.

Waste Disposal Protocol

Manganese/iodine waste to aqueous heavy-metal waste.

Viva Voce Oral Defense Questions & Answers

Q: Write the key Winkler reactions.

A: Mn2+ + 2 OH- -> Mn(OH)2; 2 Mn(OH)2 + O2 -> 2 MnO(OH)2; MnO(OH)2 + 2 I- + 4 H+ -> Mn2+ + I2 + 3 H2O.

Q: How does the Clark probe work?

A: O2 diffuses through a membrane and is reduced at a Pt cathode; the diffusion-limited current is proportional to dissolved O2.

Q: Why compare the two methods?

A: Winkler is the absolute reference method; agreement validates (calibrates) the faster probe measurement.

← Back to Experiments Directory