CHEMVEN

Potentiometric Complexometric Titration: Zinc vs EDTA with Mercury Electrode

To titrate Zn2+ with EDTA potentiometrically using a mercury indicator electrode.

Level: UniversitySafety: Advanced lab onlyTime: 90 min
Aim of Experiment: To titrate Zn2+ with EDTA potentiometrically using a mercury indicator electrode.

Theoretical Foundations

A mercury electrode in the presence of a trace of Hg-EDTA complex responds to free metal ion via HgY2- + 2e- -> Hg + Y4- coupled to Zn2+ + Y4- -> ZnY2-. The potential follows log[Zn2+] free, which plummets at the endpoint as Zn2+ is complexed, giving a sharp potential drop. (Demonstration only: mercury handling requires strict supervision.)

Required Apparatus & Chemicals

Laboratory Apparatus

  • mV meter
  • Mercury pool electrode (J-tube)
  • Reference electrode
  • Burette
  • Beaker

Chemical Reagents

  • ZnSO4 (~0.01 M)
  • EDTA standard (0.01 M)
  • Ammonia buffer pH 10
  • Trace Hg-EDTA solution
  • Mercury (instructor-handled)

Step-by-Step Practical Procedure

  1. Set up the mercury electrode under instructor supervision.
  2. Pipette 25.0 mL Zn2+ solution, add pH 10 buffer, dilute to ~75 mL.
  3. Add one drop of dilute Hg-EDTA indicator solution.
  4. Titrate with 0.01 M EDTA, recording E.
  5. Plot E and dE/dV; the endpoint is a sharp potential drop.
  6. Calculate [Zn2+].

Observations & Expected Results

E is steady near -0.9 V, then drops ~150 mV sharply at Ve.

Worked Stoichiometric Calculations

Problem: Ve = 24.6 mL of 0.0100 M EDTA for 25.0 mL sample. Find [Zn2+].

Working: Moles EDTA = 0.0246 x 0.0100 = 2.46 x 10^-4 mol = moles Zn2+. [Zn2+] = 2.46 x 10^-4/0.0250 = 9.84 x 10^-3 M.

Safety Precautions & Hazard Control

  • Mercury is toxic: instructor handles all mercury; work over a spill tray.
  • Never pour mercury down the drain.
  • Buffer at pH 10 is required for stable Zn-EDTA formation.

Waste Disposal Protocol

All mercury and mercury-contaminated solutions to dedicated mercury waste.

Viva Voce Oral Defense Questions & Answers

Q: How does the mercury electrode sense Zn2+?

A: Through the coupled equilibria HgY2-/Hg and ZnY2-/Zn2+: the electrode potential tracks free [Y4-], which is set by free [Zn2+], so it drops when Zn2+ is complexed.

Q: Why is this a demonstration-level experiment?

A: Elemental mercury is a cumulative neurotoxin; only trained staff may handle it, and many labs now substitute other indicator electrodes.

← Back to Experiments Directory