Ion-Selective Electrode: Nitrate in Soil Extracts

To determine nitrate-nitrogen in soil extracts with a nitrate ISE by the standard-addition method.

Level: UniversitySafety: SafeTime: 90 min
Aim of Experiment: To determine nitrate-nitrogen in soil extracts with a nitrate ISE by the standard-addition method.

Theoretical Foundations

The nitrate ISE (liquid membrane) follows E = const - 0.05916 log[NO3-]. Soil extracts contain interfering anions (Cl-, HCO3-), so the standard-addition (spiking) method is used: the potential shift on adding a known nitrate spike gives the original concentration without a matrix-matched calibration. For a spike doubling the volume, C_x = C_s / (10^(dE/S) - 1) x (V_s/V_x) style relations apply.

Required Apparatus & Chemicals

Laboratory Apparatus

  • Nitrate ISE + reference
  • mV meter
  • Beakers
  • Micropipette
  • Stirrer

Chemical Reagents

  • KNO3 standard (1000 mg/L NO3-)
  • Ionic strength adjuster (e.g. (NH4)2SO4 solution)
  • Soil extract (filtered)
  • Distilled water

Step-by-Step Practical Procedure

  1. Filter the soil extract and add ionic strength adjuster.
  2. Measure the steady potential E1 of 50.0 mL extract.
  3. Spike with 0.50 mL of 1000 mg/L nitrate standard; record E2.
  4. Note the electrode slope S from a quick two-standard check.
  5. Calculate the original nitrate concentration from the potential shift.
  6. Express the result as mg NO3--N per kg of soil.

Observations & Expected Results

Typical shifts are 5-15 mV for the spike; computed soil nitrate often falls in the 5-50 mg N/kg range for agricultural soils.

Worked Stoichiometric Calculations

Problem: 50.0 mL extract: E1 = 148 mV. After adding 0.50 mL of 1000 mg/L NO3- standard, E2 = 139 mV. Slope S = -58 mV/decade. Find [NO3-] in mg/L.

Working: dE = E2 - E1 = -9 mV. C_x = (C_s V_s / V_x) / (10^(-dE/S) - 1) = (1000 x 0.5/50) / (10^(9/58) - 1) = 10 / (1.428 - 1) = 23.4 mg/L NO3-.

Safety Precautions & Hazard Control

  • Keep the membrane conditioned and never let it dry out.
  • Add the same ionic strength adjuster to all solutions.
  • Chloride above ~100 mg/L interferes; note it if present.

Waste Disposal Protocol

Soil extracts and standards to drain.

Viva Voce Oral Defense Questions & Answers

Q: Why use standard addition instead of direct calibration here?

A: The soil matrix (ionic strength, interfering anions) shifts the electrode response; spiking the sample itself cancels matrix effects.

Q: Which ions interfere with the nitrate electrode?

A: Chloride, bicarbonate, nitrite and organic anions, per the selectivity coefficients of the membrane.