Ion-Selective Electrode: Fluoride Determination in Water
To determine fluoride in a water sample with a fluoride ion-selective electrode using a calibration curve.
Theoretical Foundations
The LaF3-crystal fluoride electrode follows E = constant - 0.05916 log[F-] at 25 C. TISAB (total ionic strength adjustment buffer) fixes ionic strength, buffers pH ~5-7 and complexes interfering Al3+/Fe3+ that would bind F-. A calibration plot of E vs log[F-] gives the unknown from its measured potential.
Required Apparatus & Chemicals
Laboratory Apparatus
- Fluoride ISE + reference electrode
- mV meter
- Volumetric flasks
- Beakers
- Stirrer
Chemical Reagents
- NaF standard (1000 mg/L stock)
- TISAB buffer
- Water sample
- Distilled water
Step-by-Step Practical Procedure
- Prepare fluoride standards: 0.1, 0.5, 1.0, 2.0, 5.0 mg/L.
- Add TISAB to each standard and to the sample (1:1 or per instructions).
- Measure the mV of each standard with stirring; allow stabilisation.
- Plot E versus log[F-] and check the slope (~-59 mV/decade).
- Measure the sample potential under identical conditions.
- Read [F-] from the calibration line.
Observations & Expected Results
A good electrode gives a slope of -57 to -60 mV per decade; drinking-water samples typically read 0.2-1.0 mg/L.
Worked Stoichiometric Calculations
Problem: Calibration: E = 120 - 58.5 log[F- (mg/L)]. Sample reads E = 132 mV. Find [F-].
Working: log[F-] = (120-132)/58.5 = -0.205; [F-] = 10^-0.205 = 0.62 mg/L.
Safety Precautions & Hazard Control
- Always add TISAB to standards and samples alike.
- Rinse electrodes between readings; carryover biases low-level results.
- Allow full stabilisation (1-2 min) at low concentrations.
Waste Disposal Protocol
Fluoride standards and TISAB to drain with water.
Viva Voce Oral Defense Questions & Answers
Q: What does TISAB do?
A: It holds ionic strength constant, buffers pH to 5-7 (avoiding HF formation and OH- interference), and complexes Al3+/Fe3+ that would tie up fluoride.
Q: Why is the slope negative?
A: The electrode responds to the anion F-; the Nernst equation for an anion gives E = const - 0.05916 log[F-].
Q: Why must standards and samples contain the same TISAB proportion?
A: So that ionic strength, pH and complexation conditions are identical and the calibration applies to the sample.