Potentiometric Argentometric Titration: Chloride vs AgNO3
To determine chloride potentiometrically with AgNO3 using a silver indicator electrode.
Theoretical Foundations
A silver electrode responds to Ag+ activity: E = E0(Ag+/Ag) + 0.05916 log[Ag+]. Before the endpoint [Ag+] is fixed by the AgCl solubility product ([Ag+] = Ksp/[Cl-]) and changes slowly; at the endpoint [Ag+] jumps as excess Ag+ accumulates, giving a sharp potential rise. The dE/dV peak locates Ve precisely without any indicator.
Required Apparatus & Chemicals
Laboratory Apparatus
- mV meter
- Silver indicator electrode (or silver wire)
- Calomel reference electrode
- Burette
- Beaker
- Stirrer
Chemical Reagents
- NaCl (~0.05 M)
- AgNO3 standard (0.05 M)
- Distilled water
Step-by-Step Practical Procedure
- Pipette 25.0 mL NaCl solution and dilute to ~75 mL.
- Insert electrodes, stir and record the initial potential.
- Add AgNO3 in 1 mL steps, recording E.
- Use 0.2 mL steps through the jump.
- Plot E and dE/dV versus volume.
- Read Ve from the derivative peak and calculate [Cl-].
Observations & Expected Results
E rises slowly from ~0.25 V, then jumps ~200 mV at the endpoint before levelling off.
Worked Stoichiometric Calculations
Problem: Ve = 24.8 mL of 0.0500 M AgNO3 for 25.0 mL sample. Find [Cl-].
Working: Moles Ag+ = 0.0248 x 0.0500 = 1.24 x 10^-3 mol = moles Cl-. [Cl-] = 1.24 x 10^-3/0.0250 = 0.0496 M.
Safety Precautions & Hazard Control
- Protect AgNO3 from light.
- Polish the silver electrode if the response drifts.
- Stir well to keep AgCl from coating the electrode.
Waste Disposal Protocol
Silver residues to heavy-metal waste.
Viva Voce Oral Defense Questions & Answers
Q: Why does the potential jump at the endpoint?
A: [Ag+] is suppressed by precipitation (Ksp/[Cl-]) before Ve but rises freely with excess titrant after Ve; the Nernst equation turns this concentration jump into a potential jump.
Q: Why is no indicator needed?
A: The silver electrode directly senses Ag+ activity, so the endpoint is detected instrumentally.