CHEMVEN

Conductometric Determination of Sulfate by Barium Precipitation

To determine sulfate in a sample by conductometric titration with BaCl2, following BaSO4 precipitation.

Level: UniversitySafety: Needs supervisionTime: 90 min
Aim of Experiment: To determine sulfate in a sample by conductometric titration with BaCl2, following BaSO4 precipitation.

Theoretical Foundations

Ba2+ + SO4(2-) -> BaSO4(s). Before the endpoint, sulfate is replaced by chloride (2 Cl- per SO4(2-)); conductance changes only slightly because mobilities are comparable. After the endpoint, excess Ba2+ and Cl- accumulate and conductance rises. The break point gives the sulfate content; the method avoids the tedious gravimetric BaSO4 procedure.

Required Apparatus & Chemicals

Laboratory Apparatus

  • Conductivity meter
  • Burette
  • Beaker
  • Stirrer
  • Pipette

Chemical Reagents

  • Na2SO4 sample solution (~0.05 M)
  • BaCl2 standard (0.1 M)
  • Distilled water

Step-by-Step Practical Procedure

  1. Pipette 25.0 mL sulfate sample, dilute to ~100 mL.
  2. Record conductance with stirring.
  3. Add 0.1 M BaCl2 in 1 mL steps, recording conductance.
  4. Continue past the endpoint.
  5. Plot conductance vs volume; the intersection of the two lines is Ve.
  6. Calculate sulfate molarity and mg SO4(2-) in the sample.

Observations & Expected Results

Conductance is nearly flat (slight fall) before the endpoint, then rises clearly after it; the break is sharp.

Worked Stoichiometric Calculations

Problem: Ve = 11.8 mL of 0.100 M BaCl2 for 25.0 mL sample. What mass of SO4(2-) was present? (M = 96.06)

Working: Moles Ba2+ = 0.0118 x 0.100 = 1.18 x 10^-3 mol = moles SO4(2-). Mass = 1.18 x 10^-3 x 96.06 = 0.113 g.

Safety Precautions & Hazard Control

  • Stir well; BaSO4 can coat electrodes and dull the response.
  • Keep temperature constant.
  • Standardise BaCl2 if high accuracy is needed.

Waste Disposal Protocol

BaSO4 precipitate to heavy-metal solid waste; filtrate to drain.

Viva Voce Oral Defense Questions & Answers

Q: Why is the pre-endpoint conductance nearly flat?

A: One SO4(2-) is replaced by two Cl- of similar total mobility, so the ionic conductance barely changes.

Q: Why is this faster than gravimetry?

A: No filtration, ignition or weighing of BaSO4 is needed; the endpoint comes directly from the titration curve.

← Back to Experiments Directory