HPLC Concept: Reversed-Phase Separation of Caffeine and Aspirin
To separate caffeine and aspirin by reversed-phase HPLC and quantify caffeine in a soft drink by external calibration.
Theoretical Foundations
Reversed-phase HPLC uses a non-polar C18 stationary phase with a polar mobile phase (water-methanol). Retention follows hydrophobicity: the more polar analyte elutes first. Aspirin (partly ionised at neutral pH) elutes before the less polar caffeine under acidic conditions that suppress ionisation. External calibration with standard peak areas converts the sample peak area to concentration.
Required Apparatus & Chemicals
Laboratory Apparatus
- HPLC with UV detector and C18 column
- Autosampler vials
- 0.45 um syringe filters
- Volumetric flasks
Chemical Reagents
- Caffeine standard
- Aspirin standard
- Methanol (HPLC grade)
- Phosphoric acid / phosphate buffer pH 3
- Degassed soft drink sample
Step-by-Step Practical Procedure
- Prepare a mobile phase of methanol:buffer (pH 3) and degas it.
- Equilibrate the C18 column at 1 mL/min until the baseline is stable.
- Inject caffeine and aspirin standards separately to record retention times at 254 nm.
- Inject a series of caffeine standards (5-50 mg/L) and build a peak-area calibration curve.
- Filter the degassed soft drink and inject it.
- Identify the caffeine peak by retention time and read its concentration from the calibration curve.
Observations & Expected Results
Aspirin elutes first (more polar at pH 3, partly retained), caffeine second; typical retention times are 3-4 min and 5-6 min at 1 mL/min on a 15 cm column.
Worked Stoichiometric Calculations
Problem: Calibration: area = 1250 x C (mg/L). The drink gives caffeine area 25000. What is its caffeine content?
Working: C = 25000/1250 = 20 mg/L. (Compare with typical cola values of ~100 mg/L; the sample may have been diluted.)
Safety Precautions & Hazard Control
- Never run the pump dry; always keep the inlet in mobile phase.
- Filter all samples through 0.45 um membranes to protect the column.
- High-pressure fittings must be tightened by trained staff only.
Waste Disposal Protocol
Methanol-containing mobile phase and vials to halogen-free organic waste.
Viva Voce Oral Defense Questions & Answers
Q: Why is the mobile phase acidic for this separation?
A: Low pH suppresses ionisation of aspirin's carboxylic acid group, making it neutral and retained; at high pH it would ionise and elute in the void.
Q: Why does caffeine elute after aspirin in reversed phase?
A: Caffeine is less polar/more hydrophobic, so it partitions more strongly into the C18 stationary phase.
Q: What is external calibration?
A: Quantifying the sample by comparing its peak area to a calibration curve prepared from separate standard solutions.