ANALYTICALLevel: BeginnerAcademic Reference

Overview & Foundations

A burette is best thought of as a counting device, not as a tap. The analyst fills it with a solution of accurately known concentration, the titrant, and adds it drop by drop to a measured volume of the solution being tested. Each drop delivers a known number of moles, so the volume used is really a mole count in disguise. Two points must never be confused. The equivalence point is the true stoichiometric point, where the moles added exactly match the moles present according to the equation. The endpoint is what the analyst actually sees: the moment an indicator changes colour. Good technique makes those two points coincide, by choosing the right indicator, adding titrant slowly near the end, and repeating until volumes agree closely.

There are four families of titration, classified by the reaction used. Acid–base titrations use neutralisation. Redox titrations use electron transfer, for example permanganate or iodine oxidising the analyte. Complexometric titrations use complex formation, above all EDTA binding metal ions. Precipitation titrations use the formation of an insoluble solid, classically silver nitrate precipitating chloride. The family decides everything else: what indicator can work, what side reactions threaten accuracy, and what the titration curve looks like. Asking which family it is guides every later choice.

Acid–base curves show why indicator choice matters. When a strong acid is titrated with a strong base, the pH stays low, then jumps steeply through neutrality, then levels off high; the equivalence point is at pH 7, and any indicator changing inside that jump gives the right volume. Titrate a weak acid with a strong base and the picture shifts: at equivalence the solution contains the conjugate base, which hydrolyses, so equivalence lies above pH 7 and the steep section is shorter and higher. Titrate a weak base with a strong acid and equivalence lies below pH 7, for the mirror-image reason. The half-equivalence point has its own gift: halfway there, acid and conjugate base are present in equal amounts, and the pH equals the pKa — the simplest accurate way to measure one.

An indicator is itself a weak acid or base whose two forms have different colours, so it changes over a range of about two pH units, not at a single pH. The rule is simple and often broken: choose the indicator whose range sits on the steep, near-vertical part of the curve. Phenolphthalein, changing between pH 8.2 and 10.0, fits a weak acid titrated with a strong base. Methyl orange, changing between pH 3.1 and 4.4, fits a weak base titrated with a strong acid. Bromothymol blue, changing between pH 6.0 and 7.6, suits endpoints near neutrality. Litmus changes so gradually, over roughly pH 5 to 8, that it can never mark a precise endpoint. Using methyl orange for a weak acid–strong base titration does not give a slightly wrong answer; the colour changes well before equivalence.

Redox titrations often manage without a conventional indicator. Potassium manganate(VII) is intensely purple as MnO₄⁻ and almost colourless as Mn²⁺, so in acid the titrant bleaches itself while any reducing agent remains; the first permanent pale pink, from a slight excess of MnO₄⁻, is the endpoint, and the titration is self-indicating. Iodometric titrations use starch instead: starch forms an intense blue-black complex with iodine, and the endpoint is the disappearance of that blue colour, with starch added only near the end so the complex does not release its iodine too slowly. Conditions matter as much as volumes: permanganate titrations need acid solution, or a brown precipitate of manganese dioxide forms and the stoichiometry changes.

Complexometric and precipitation titrations extend the counting idea to ions that are neither acids nor oxidisers. EDTA wraps around a metal ion and binds it in a strict 1:1 ratio, whatever the ion's charge, which makes the arithmetic unusually clean. With Eriochrome Black T as indicator, EDTA titration measures the calcium and magnesium in water and so gives water hardness. Precipitation titration is older: in the Mohr method, silver nitrate is titrated against chloride, silver chloride precipitates first, and only when the chloride is exhausted does silver react with a chromate indicator to give a red precipitate — the work of Mohr, whose 1855 textbook founded modern titrimetry. Where no clean direct endpoint exists, the analyst adds a known excess of one reagent and back-titrates what is left.

Accuracy is won by small habits repeated every time. The burette is read at the bottom of the meniscus, at eye level. A rough titration first finds the endpoint; careful titrations then approach it drop by drop, finally by fractions of a drop washed in with distilled water, which changes no mole count. Concordant results are averaged; a single titration proves nothing, because one overshoot is invisible without a second attempt. Rinsing matters too: burette and pipette are rinsed with the solutions they will hold, while the conical flask is rinsed only with distilled water, since extra water changes no moles, but stray analyte solution would be counted twice.

Core Chemical Concepts

Governing Formulas & Laws

n = c × V
Moles from concentration in mol/L and volume in litres — the burette counts moles
c₁V₁ = c₂V₂ (for 1:1 reactions)
At equivalence the moles of titrant equal the moles of analyte
pH = pKa at half-equivalence
Half the acid is converted, so acid and conjugate base are equal in amount
EDTA : metal ion = 1 : 1
One EDTA chelates one metal ion, whatever the ion's charge

Worked Numerical Examples

25.0 mL of sodium hydroxide solution needs 20.0 mL of 0.100 mol/L HCl to reach the endpoint. Find the NaOH concentration.
First count the moles of acid used: n(HCl) = c × V = 0.100 mol/L × 0.0200 L = 2.00×10⁻³ mol. The reaction HCl + NaOH → NaCl + H₂O is 1:1, so n(NaOH) = 2.00×10⁻³ mol as well. That amount was in 25.0 mL, so c(NaOH) = n / V = 2.00×10⁻³ / 0.0250 = 0.0800 mol/L.
Why is potassium manganate(VII) titration done without an indicator, and what is the endpoint?
No indicator is needed because the titrant signals its own endpoint. MnO₄⁻ is intensely purple, while Mn²⁺ in acid is almost colourless, so every drop is decolourised while reducing agent remains. Once it is exhausted, the next drop's colour survives. The endpoint is the first permanent pale pink — the titration is self-indicating.

Industrial & Laboratory Applications

Common Student Misconceptions

Confusing the endpoint with the equivalence point and assuming any indicator will do
Choosing an indicator whose range misses the steep section, for example methyl orange for a weak acid–strong base titration
Misreading the burette scale, or overshooting the endpoint by adding titrant in a stream

Everyday Life & Real-World Relevance

Practice & Review Questions

Endpoint or equivalence point — which one does the calculation need?
The calculation needs the equivalence point, where moles match the equation exactly. The analyst can only observe the endpoint, so the indicator and technique are chosen to make the endpoint fall as close to equivalence as possible.
Why is litmus a poor titration indicator?
Litmus changes colour gradually over a broad range, roughly pH 5 to 8, with no sharp moment of change. It shows acidity or alkalinity, but cannot pin down the single drop at which equivalence is reached.
What is a back titration, and when is it used?
A known excess of one reagent is added and allowed to react completely; the leftover reagent is then titrated. It is used when the direct reaction is slow, the analyte is insoluble, or no direct endpoint exists.