Thermal Analysis: TGA, DSC, DTA

Heat it and watch: mass loss, heat flow and transitions reveal composition and stability.

ANALYTICALLevel: IntermediateAcademic Reference

Overview & Foundations

TGA (thermogravimetric analysis): mass vs temperature — steps reveal decomposition, water loss (hydrates), filler content in polymers, and thermal stability rankings. Derivative (DTG) sharpens overlapping steps.

DSC (differential scanning calorimetry): heat flow vs temperature — measures Tg (glass transition, step), crystallization (exotherm), melting (endotherm), and heat capacity. Purity from melting-point depression; polymer crystallinity from ΔH_fusion.

DTA (differential thermal analysis): temperature difference vs reference — the older cousin of DSC, still used for high-temperature ceramics and phase diagrams. Combined TGA-DSC runs both at once. Evolved-gas analysis (TGA-MS/FTIR) identifies what leaves at each step.

Core Chemical Concepts

Governing Formulas & Laws

% mass loss ⇒ stoichiometry
Hydrate formulas from TGA

Worked Numerical Examples

Q: TGA shows 36% mass loss for CuSO₄·xH₂O leaving CuSO₄. x?
A: 0.36 = 18x/(159.6+18x) → x ≈ 5. Pentahydrate confirmed.

Industrial & Laboratory Applications

Common Student Misconceptions

Heating rate too fast (steps merge)
Wrong atmosphere (oxidation vs inert)

Everyday Life & Real-World Relevance

Practice & Review Questions

DSC vs DTA?
DSC measures heat flow directly (quantitative enthalpies); DTA measures temperature difference (qualitative, simpler, higher T).
How does TGA find hydrate formulas?
Each mass-loss step is water leaving — the % loss gives x in the formula.

Scholarly Literature & References